Biomechanics of two external fixator devices used in rat femoral fractures

Osagie-Clouard, L; Kaufmann, J; Blunn, G; Coathup, M; Pendegrass, C; Meeson, R; Briggs, T; (2019) Biomechanics of two external fixator devices used in rat femoral fractures. Journal of Orthopaedic Research , 37 (2) pp. 293-298. 10.1002/jor.24034. Green open access

[thumbnail of Osagie-Clouard 2018 Biomechanics of Two External Fixator Devices Used in Rat Femoral Fractures-accepted version.pdf]
Preview
Text
Osagie-Clouard 2018 Biomechanics of Two External Fixator Devices Used in Rat Femoral Fractures-accepted version.pdf - Accepted Version

Download (780kB) | Preview

Abstract

The use of external fixators allows for the direct investigation of newly formed interfragmentary bone, and the radiographic evaluation of the fracture. We validated the results of a finite element model with the in vitro stiffness' of two widely used external fixator devices used for in vivo analysis of fracture healing in rat femoral fractures with differing construction (Ti alloy ExFix1 and PEEK ExFix2). Rat femoral fracture fixation was modelled using two external fixators. For both constructs an osteotomy of 2.75 mm was used, and offset maintained at 5 mm. Tufnol, served as standardized substitutes for rat femora. Constructs were loaded under axial compression and torsion. Overall axial and torsional stiffness were compared between the in vitro models and FE results. FE models were also used to compare the fracture movement and overall pattern of von Mises stress across the external fixators. In vitro axial stiffness of ExFix1 was 29.26 N/mm ± 3.83 compared to ExFix2 6.31 N/mm ± 0.67 (p* < 0.05). Torsional stiffness of ExFix1 was 47.5 Nmm/° ± 2.71 compared to ExFix2 at 19.1 Nmm/° ± 1.18 (p* < 0.05). FE results predicted similar comparative ratios between the ExFix1 and 2 as the in vitro studies. FE results predicted considerably larger interfragmentary motion in the ExFix2 comparing to ExFix1. We demonstrated significant differences in the stiffness' of the two external fixators as one would expect from such variable designs; yet, importantly we validated the utility of an FE model for the analysis and prediction of changes in fracture mechanics dependent on fixator choice. This article is protected by copyright. All rights reserved.

Type: Article
Title: Biomechanics of two external fixator devices used in rat femoral fractures
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/jor.24034
Publisher version: http://dx.doi.org/10.1002/jor.24034
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: finite element analysis, biomechanics, fracture fixation
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci > Department of Ortho and MSK Science
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering
URI: https://discovery.ucl.ac.uk/id/eprint/10048295
Downloads since deposit
52Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item