Effect of cochlear implant electrode insertion on middle-ear function as measured by intra-operative laser Doppler vibrometry

Donnelly, N; Bibas, A; Jiang, D; Bamiou, DE; Santulli, C; Jeronimidis, G; O'Connor, AF; (2009) Effect of cochlear implant electrode insertion on middle-ear function as measured by intra-operative laser Doppler vibrometry. Journal of Laryngology and Otology , 123 (7) 723 - 729. 10.1017/S0022215109004290. Green open access

[thumbnail of S0022215109004290.pdf]
Preview
PDF
S0022215109004290.pdf
Available under License : See the attached licence file.

Download (719kB)

Abstract

Hypothesis: The aim of this study was to investigate the impact of cochlear implant electrode insertion on middle-ear low frequency function in humans.Background: Preservation of residual low frequency hearing with addition of electrical speech processing can improve the speech perception abilities and hearing in noise of cochlear implant users. Preservation of low frequency hearing requires an intact middle-ear conductive mechanism in addition to intact inner-ear mechanisms. Little is known about the effect of a cochlear implant electrode on middle-ear function.Methods: Stapes displacement was measured in seven patients undergoing cochlear implantation. Measurements were carried out intra-operatively before and after electrode insertion. Each patient acted as his or her own control. Sound was delivered into the external auditory canal via a speaker and calibrated via a probe microphone. The speaker and probe microphone were integrated into an individually custom-made ear mould. Ossicular displacement in response to a multisine stimulus at 80 dB SPL was measured at the incudostapedial joint via the posterior tympanotomy, using an operating microscope mounted laser Doppler vibrometry system.Results: Insertion of a cochlear implant electrode into the scala tympani had a variable effect on stapes displacement. In three patients, there was little change in stapes displacement following electrode insertion. In two patients, there was a significant increase, while in a further two there was a significant reduction in stapes displacement. This variability may reflect alteration of cochlear impedance, possibly due to differing loss of perilymph associated with the electrode insertion.Conclusion: Insertion of a cochlear implant electrode produces a change in stapes displacement at low frequencies, which may have an effect on residual low frequency hearing thresholds.

Type: Article
Title: Effect of cochlear implant electrode insertion on middle-ear function as measured by intra-operative laser Doppler vibrometry
Open access status: An open access version is available from UCL Discovery
DOI: 10.1017/S0022215109004290
Publisher version: http://dx.doi.org/10.1017/S0022215109004290
Language: English
Additional information: Copyright Cambridge University Press 2009
Keywords: Cochlear implant, Middle ear, Physiology, Laser doppler vibrometry, Acoustic stimulation, Auditory-system, Hearing, Preservation
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > The Ear Institute
URI: https://discovery.ucl.ac.uk/id/eprint/54754
Downloads since deposit
284Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item