Saeidi, N and Demosthenous, A and Donaldson, N and Alderman, J (2009) Design and Fabrication of Corrosion and Humidity Sensors for Performance Evaluation of Chip Scale Hermetic Packages for Biomedical Implantable Devices. In: 2009 EUROPEAN MICROELECTRONICS AND PACKAGING CONFERENCE (EMPC 2009), VOLS 1 AND 2. (pp. 447 - 450). IEEE
Full text not available from this repository.
Abstract
In this paper we report on the development of a set of test chips that can be used to verify the sealing techniques with biomedical implants as their tat-get application. These test chips have been designed and fabricated based on the Tyndall's in-house CMOS 1.5 micron technology. Each chip includes humidity, temperature and/or corrosion sensors. Different designs and sizes have been considered for these sensors to compare the sensors sensitivity and to select the optimum ones for the final chip. The sensors are being tested and the results obtained for two of the humidity sensors are presented.
| Type: | Proceedings paper |
|---|---|
| Title: | Design and Fabrication of Corrosion and Humidity Sensors for Performance Evaluation of Chip Scale Hermetic Packages for Biomedical Implantable Devices |
| Event: | European Microelectronics and Packaging Conference (EMPC 2009) |
| Location: | Rimini, ITALY |
| Dates: | 2009-06-16 - 2009-06-18 |
| ISBN-13: | 978-1-4244-4722-0 |
| Keywords: | Biomedical Packaging, Hermeticity Testing, Test Chip, Capacitive Humidity Sensors |
| UCL classification: | UCL > School of BEAMS > Faculty of Engineering Science > Electronic and Electrical Engineering UCL > School of BEAMS > Faculty of Engineering Science > Medical Physics and Bioengineering |
Archive Staff Only: edit this record