UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

3D Real-Time Echocardiography Combined with Mini Pressure Wire Generate Reliable Pressure-Volume Loops in Small Hearts

Herberg, U; Linden, K; Dewald, O; Gatzweiler, E; Seehase, M; Duerr, GD; Doerner, J; ... Breuer, J; + view all (2016) 3D Real-Time Echocardiography Combined with Mini Pressure Wire Generate Reliable Pressure-Volume Loops in Small Hearts. PLoS One , 11 (10) , Article e0165397. 10.1371/journal.pone.0165397. Green open access

[thumbnail of Breuer_journal.pone.0165397.PDF]
Preview
Text
Breuer_journal.pone.0165397.PDF - Published Version

Download (2MB) | Preview

Abstract

BACKGROUND: Pressure-volume loops (PVL) provide vital information regarding ventricular performance and pathophysiology in cardiac disease. Unfortunately, acquisition of PVL by conductance technology is not feasible in neonates and small children due to the available human catheter size and resulting invasiveness. The aim of the study was to validate the accuracy of PVL in small hearts using volume data obtained by real-time three-dimensional echocardiography (3DE) and simultaneously acquired pressure data. METHODS: In 17 piglets (weight range: 3.6–8.0 kg) left ventricular PVL were generated by 3DE and simultaneous recordings of ventricular pressure using a mini pressure wire (PVL3D). PVL3D were compared to conductance catheter measurements (PVLCond) under various hemodynamic conditions (baseline, alpha-adrenergic stimulation with phenylephrine, beta-adrenoreceptor-blockage using esmolol). In order to validate the accuracy of 3D volumetric data, cardiac magnetic resonance imaging (CMR) was performed in another 8 piglets. RESULTS: Correlation between CMR- and 3DE-derived volumes was good (enddiastolic volume: mean bias -0.03ml ±1.34ml). Computation of PVL3D in small hearts was feasible and comparable to results obtained by conductance technology. Bland-Altman analysis showed a low bias between PVL3D and PVLCond. Systolic and diastolic parameters were closely associated (Intraclass-Correlation Coefficient for: systolic myocardial elastance 0.95, arterial elastance 0.93, diastolic relaxation constant tau 0.90, indexed end-diastolic volume 0.98). Hemodynamic changes under different conditions were well detected by both methods (ICC 0.82 to 0.98). Inter- and intra-observer coefficients of variation were below 5% for all parameters. CONCLUSIONS: PVL3D generated from 3DE combined with mini pressure wire represent a novel, feasible and reliable method to assess different hemodynamic conditions of cardiac function in hearts comparable to neonate and infant size. This methodology may be integrated into clinical practice and cardiac catheterization programs and has the capability to contribute to clinical decision making even in small hearts.

Type: Article
Title: 3D Real-Time Echocardiography Combined with Mini Pressure Wire Generate Reliable Pressure-Volume Loops in Small Hearts
Open access status: An open access version is available from UCL Discovery
DOI: 10.1371/journal.pone.0165397
Publisher version: http://doi.org/10.1371/journal.pone.0165397
Language: English
Additional information: © 2016 Herberg et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Keywords: Science & Technology, Multidisciplinary Sciences, Science & Technology - Other Topics, RIGHT-VENTRICULAR FUNCTION, SINGLE-BEAT ESTIMATION, 3-DIMENSIONAL ECHOCARDIOGRAPHY, CONDUCTANCE CATHETER, VASCULAR-RESISTANCE, FEASIBILITY, VALIDATION, INFANTS, DISEASE, QUANTIFICATION
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > UCL GOS Institute of Child Health
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > UCL GOS Institute of Child Health > Infection, Immunity and Inflammation Dept
URI: https://discovery.ucl.ac.uk/id/eprint/1553691
Downloads since deposit
60Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item