Ashworth, PC and Pickwell-MacPherson, E and Provenzano, E and Pinder, SE and Purushotham, AD and Pepper, M and Wallace, VP (2009) Terahertz pulsed spectroscopy of freshly excised human breast cancer. OPT EXPRESS , 17 (15) 12444 - 12454.
Abstract
The complex refractive indices of freshly excised healthy breast tissue and breast cancers collected from 20 patients were measured in the range of 0.15 - 2.0 THz using a portable terahertz pulsed transmission spectrometer. Histology was performed to classify the tissue samples as healthy adipose tissue, healthy fibrous breast tissue, or breast cancers. The average complex refractive index was determined for each group and it was found that samples containing cancer had a higher refractive index and absorption coefficient. The terahertz properties of the tissues were also used to simulate the impulse response functions expected when imaging breast tissue in a reflection geometry as in terahertz pulsed imaging (TPI). Our results indicate that both TPS and TPI can be used to distinguish between healthy adipose breast tissue, healthy fibrous breast tissue and breast cancer due to the differences in the fundamental optical properties. (C) 2009 Optical Society of America
| Type: | Article |
|---|---|
| Title: | Terahertz pulsed spectroscopy of freshly excised human breast cancer |
| Open access status: | An open access publication |
| Publisher version: | http://www.opticsexpress.org |
| Keywords: | BASAL-CELL CARCINOMA |
| UCL classification: | UCL > School of BEAMS > Faculty of Engineering Science > Electronic and Electrical Engineering |