Performance of the platelet function analyser PFA-100 (R) in testing abnormalities of primary haemostasis

Harrison, P; Robinson, MSC; Mackie, IJ; Joseph, J; McDonald, SJ; Liesner, R; Savidge, GF; ... Machin, SJ; + view all (1999) Performance of the platelet function analyser PFA-100 (R) in testing abnormalities of primary haemostasis. BLOOD COAGUL FIBRIN , 10 (1) 25 - 31.

Full text not available from this repository.

Abstract

The PFA-100(R) device is a new instrument for the in-vitro testing of platelet function. Primary haemostasis is stimulated by recording the closure time taken for platelets to seal a 150 mu m aperture in the centre of a membrane coated with collagen and either epinephrine or ADP, Patients with type 3 von Willebrand's disease (n = 4) all had infinitely prolonged closure times (> 200 s) with both types of cartridge. A patient with afibrinogenemia exhibited only slightly prolonged closure times of 111 and 166 s for the ADP and epinephrine membranes, respectively. Patients with Glanzmann's thrombasthenia (n = 6) and Bernard Soulier syndrome (n = 2) had grossly prolonged closure times (> 200 s) with both types of cartridges. These results confirmed that the PFA-100(R) system was highly dependent on normal von Willebrand factor, glycoprotein Ib and glycoprotein IIb/IIIa levels but not on plasma fibrinogen. Patients with storage pool disease (n = 6) and Hermansky Pudlak syndrome (n = 7) had prolonged closure times with the epinephrine cartridge. There was no evidence of enhanced platelet function in patients with antiphospholipid syndrome, in sickle-cell disease or thalassemia. However, ingestion of aspirin resulted in a near consistent and significant prolongation of the closure time for the epinephrine cartridge but not for the ADP cartridge in both normal subjects and patients. The test offers a reliable, reproducible, rapid and simple means of assessing high-shear platelet function in vitro. (C) 1999 Lippincott Williams & Wilkins.

Type: Article
Title: Performance of the platelet function analyser PFA-100 (R) in testing abnormalities of primary haemostasis
Keywords: platelets, bleeding time, PFA-100 (R), platelet disorders, von Willebrand's disease, FUNCTION ANALYZER, BLEEDING-TIME, SYSTEM
UCL classification: UCL > Provost and Vice Provost Offices
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Cancer Institute
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Cancer Institute > Research Department of Haematology
URI: http://discovery.ucl.ac.uk/id/eprint/124291
Downloads since deposit
0Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item