UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Sizing the aortic annulus with a robotised, commercially available soft balloon catheter: in vitro study on idealised phantoms

Palombi, A; Bosi, G; Di Giuseppe, S; De Momi, E; Homer-Vanniasinkam, S; Burriesci, G; Wurdemann, H; (2019) Sizing the aortic annulus with a robotised, commercially available soft balloon catheter: in vitro study on idealised phantoms. In: Proceedings of 2019 International Conference on Robotics and Automation (ICRA). IEEE: Montreal, QC, Canada, Canada. Green open access

[thumbnail of a__palombi___sizing_the_aortic_annulus_with_a_soft_robotic_balloon_catheter__in_vitrostudy_on_idealised_phantoms___iros_2018-1.pdf]
Preview
Text
a__palombi___sizing_the_aortic_annulus_with_a_soft_robotic_balloon_catheter__in_vitrostudy_on_idealised_phantoms___iros_2018-1.pdf - Accepted Version

Download (3MB) | Preview

Abstract

Transcatheter aortic valve implantation (TAVI) is a minimally invasive surgical technique to treat aortic heart valve diseases. According to current clinical guidelines, the implanted prosthetic valve replacing the native one is selected based on pre-operative size assessment of the aortic annulus through different imaging techniques. That very often leads to suboptimal device selection resulting in major complications, such as prosthetic valve leakage or interruption of the cardiac electrical signal. In this paper, we propose a new, intra-operative approach to determine the diameter of theaortic annulus exploiting intra-balloon pressure and volume data, acquired from a robotised valvuloplasty balloon catheter. An inflation device, capable of collecting real-time intra-balloon pressure and volume data, was designed and interfaced with a commercially available valvuloplasty balloon catheter. A sizing algorithm allowing to precisely estimate the annular diameter was integrated. The algorithm relies on a characterised analytical model of the balloon free inflation and an iterative method based on linear regression. In vitro tests were performed on idealised aortic phantoms. Experimental results show that pressure-volume data can be used to determine annular diameters bigger than the unstretched diameter of the balloon catheter. For these cases, the proposed approach exhibited good precision (maximum average error 0.93%) and good repeatability (maximum standard deviation ±0.11 mm).

Type: Proceedings paper
Title: Sizing the aortic annulus with a robotised, commercially available soft balloon catheter: in vitro study on idealised phantoms
Event: IEEE International Conference on Robotics and Automation
Location: Montreal
Dates: 19 May 2019 - 24 May 2019
Open access status: An open access version is available from UCL Discovery
DOI: 10.1109/ICRA.2019.8793894
Publisher version: https://doi.org/10.1109/ICRA.2019.8793894
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering
URI: https://discovery.ucl.ac.uk/id/eprint/10075707