Peruzzo, P;
Susin, FM;
Colli, A;
Burriesci, G;
(2019)
In vitro assessment of pacing as therapy for aortic regurgitation.
Open Heart
, 6
(1)
, Article e000976. 10.1136/openhrt-2018-000976.
Preview |
Text
e000976.full.pdf - Published Version Download (962kB) | Preview |
Abstract
Background and objective: Clinical evaluation of pacing therapy in mitigating the aortic insufficiency after transchateter aortic valve implantation often gives contradictory outcomes. This study presents an in vitro investigation aimed at clarifying the effect of pacing on paravalvular leakage. / Methods: A series of in vitro tests reproducing the heart operating changes clinically obtained by pacing was carried out in a 26 mm Edwards Sapien XT prosthesis with mild paravalvular leakage. The effect of pacing on the regurgitant volumes per cycle and per minute was quantified, and the energy and power consumed by the left ventricle were calculated. / Results: Results indicate that though pacing results in some reduction in the total regurgitation per cycle, the volume of fluid regurgitating per minute increases substantially, causing overload of left ventricle. / Conclusions: Our tests indicate no effective haemodynamic benefit from pacing, suggesting a prudential clinical use of this therapy for the treatment of postoperative aortic regurgitation.
Type: | Article |
---|---|
Title: | In vitro assessment of pacing as therapy for aortic regurgitation |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1136/openhrt-2018-000976 |
Publisher version: | https://doi.org/10.1136/openhrt-2018-000976 |
Language: | English |
Additional information: | This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/. |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering |
URI: | https://discovery.ucl.ac.uk/id/eprint/10074891 |
Archive Staff Only
View Item |