A Case Report of Complete Blockage of a Baerveldt Glaucoma Implant Following Insertion of a 3-0 Supramid Suture

Lee, RMH; Bouremel, Y; Eames, I; Brocchini, S; Khaw, PT; (2019) A Case Report of Complete Blockage of a Baerveldt Glaucoma Implant Following Insertion of a 3-0 Supramid Suture. Journal of Glaucoma , 28 (5) e75-e76. 10.1097/IJG.0000000000001198. Green open access

[thumbnail of 2018 12 17 BGI Balloon Case Report v4.pdf]
Preview
Text
2018 12 17 BGI Balloon Case Report v4.pdf - Accepted Version

Download (146kB) | Preview

Abstract

PURPOSE: The aim of this study was to present a case of a Baerveldt glaucoma implant lumen being completely occluded with a 3-0 Supramid stent suture. PATIENT AND METHODS: The patient underwent Baerveldt glaucoma implant surgery with placement of an intraluminal 3-0 Supramid stent suture that acts to restrict flow across the device and reduce the risk of postoperative hypotony. Following suturing of the implant to the sclera, the device was flow tested. No flow was observed through the device tube and a significant ballooning of the tube diameter occurred with increased pressure on the device. The device was explanted from the eye and replaced with a different implant without further postoperative complication. The explanted device was assessed using custom microfluidic equipment in an in vitro environment. RESULTS: This phenomenon occurred despite using several different batches of the 3-0 Supramid stent suture and the device had to be removed and replaced with another device without complication. In vitro microfluidic assessment of the device demonstrated no flow across the device tube despite over 150 mm Hg of pressure being exerted on the device. CONCLUSIONS: We hypothesize that the blockage occurred at the junction between the device tube and plate and that the ballooning phenomenon observed was due to a defect in the tube wall. This case highlights the importance of flow testing all glaucoma drainage devices before insertion given the variation in manufacturing conditions to avoid the risk of intraoperative complications.

Type: Article
Title: A Case Report of Complete Blockage of a Baerveldt Glaucoma Implant Following Insertion of a 3-0 Supramid Suture
Open access status: An open access version is available from UCL Discovery
DOI: 10.1097/IJG.0000000000001198
Publisher version: https://doi.org/10.1097/IJG.0000000000001198
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > Institute of Ophthalmology
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Life Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Life Sciences > UCL School of Pharmacy
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Life Sciences > UCL School of Pharmacy > Pharmaceutics
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering
URI: https://discovery.ucl.ac.uk/id/eprint/10066546
Downloads since deposit
135Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item