UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

The clinical relevance of distortion correction in presurgical fMRI at 7 T

Lima Cardoso, P; Dymerska, B; Bachratá, B; Fischmeister, FPS; Mahr, N; Matt, E; Trattnig, S; ... Robinson, SD; + view all (2018) The clinical relevance of distortion correction in presurgical fMRI at 7 T. NeuroImage , 168 pp. 490-498. 10.1016/j.neuroimage.2016.12.070. Green open access

[thumbnail of Cardoso_-s2.0-S1053811916307972-main.pdf]
Preview
Text
Cardoso_-s2.0-S1053811916307972-main.pdf - Published Version

Download (1MB) | Preview

Abstract

Presurgical planning with fMRI benefits from increased reliability and the possibility to reduce measurement time introduced by using ultra-high field. Echo-planar imaging suffers, however, from geometric distortions which scale with field strength and potentially give rise to clinically significant displacement of functional activation. We evaluate the effectiveness of a dynamic distortion correction (DDC) method based on unmodified single-echo EPI in the context of simulated presurgical planning fMRI at 7 T and compare it with static distortion correction (SDC). The extent of distortion in EPI and activation shifts are investigated in a group of eleven patients with a range of neuropathologies who performed a motor task. The consequences of neglecting to correct images for susceptibility-induced distortions are assessed in a clinical context. It was possible to generate time series of EPI-based field maps which were free of artifacts in the eloquent brain areas relevant to presurgical fMRI, despite the presence of signal dropouts caused by pathologies and post-operative sites. Distortions of up to 5.1 mm were observed in the primary motor cortex in raw EPI. These were accurately corrected with DDC and slightly less accurately with SDC. The dynamic nature of distortions in UHF clinical fMRI was demonstrated via investigation of temporal variation in voxel shift maps, confirming the potential inadequacy of SDC based on a single reference field map, particularly in the vicinity of pathologies or in the presence of motion. In two patients, the distortion correction was potentially clinically significant in that it might have affected the localization or interpretation of activation and could thereby have influenced the treatment plan. Distortion correction is shown to be effective and clinically relevant in presurgical planning at 7 T.

Type: Article
Title: The clinical relevance of distortion correction in presurgical fMRI at 7 T
Open access status: An open access version is available from UCL Discovery
DOI: 10.1016/j.neuroimage.2016.12.070
Publisher version: http://dx.doi.org/10.1016/j.neuroimage.2016.12.070
Language: English
Additional information: © 2016 The Authors. Published by Elsevier Inc. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/BY-NC-ND/4.0/).
Keywords: Dynamic distortion correction, fMRI, Presurgical planning, Ultra-high field, Patients, Moto
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology > Imaging Neuroscience
URI: https://discovery.ucl.ac.uk/id/eprint/10059603
Downloads since deposit
81Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item