Propensity score matched comparison of subcutaneous and transvenous implantable cardioverter-defibrillator therapy in the SIMPLE and EFFORTLESS studies

Brouwer, TF; Knops, RE; Kutyifa, V; Barr, C; Mondésert, B; Boersma, LVA; Lambiase, PD; ... Healey, JS; + view all (2018) Propensity score matched comparison of subcutaneous and transvenous implantable cardioverter-defibrillator therapy in the SIMPLE and EFFORTLESS studies. EP Europace , 20 (FI2) f240-f248. 10.1093/europace/euy083. Green open access

[thumbnail of Lambiase_Manuscript document R1 clean.pdf]
Preview
Text
Lambiase_Manuscript document R1 clean.pdf - Accepted Version

Download (704kB) | Preview
[thumbnail of Lambiase_Supplemental_material.pdf]
Preview
Text
Lambiase_Supplemental_material.pdf - Accepted Version

Download (149kB) | Preview

Abstract

Aims: Comparison of outcomes between subcutaneous and transvenous implantable cardioverter-defibrillator (S-ICD and TV-ICD) therapy is hampered by varying patient characteristics and complication definitions. The aim of this analysis is to compare clinical outcomes of S-ICD and TV-ICD therapy in a matched cohort. // Methods and results: Patients implanted with de novo implantable cardioverter-defibrillators without need for pacing were selected from two studies: SIMPLE (n = 1091 single and n = 553 dual chamber TV-ICDs) and EFFORTLESS (n = 798 S-ICDs). Subcutaneous implantable cardioverter-defibrillator patients were 1:1 matched on propensity score to TV-ICD patients. Propensity scores were calculated using 15 baseline characteristics including diagnosis. The Kaplan–Meier estimates for complications requiring invasive intervention, appropriate shocks, and inappropriate shocks were calculated at 3 years follow-up. The primary analysis yielded 391 patients pairs with balanced baseline characteristics, with mean age 55 ± 14 years, 49% ischaemic cardiomyopathy, mean left ventricular ejection fraction 40%, 71% primary prevention, and 89% of TV-ICDs were single chamber. Follow-up was mean 2.9 years in the S-ICD arm vs. 3.3 in the TV-ICD arm. All-cause complications occurred in 9.0% of S-ICD vs. 6.5% of TV-ICD patients, P = 0.29. Appropriate shocks occurred in 9.9% of S-ICD vs. 13.8% in TV-ICD patients, P = 0.03 and inappropriate shocks in 11.9% in S-ICD vs. 8.9% in TV-ICD patients (P = 0.07). Total shock burden (20 vs. 31, P = 0.05) and appropriate shock burden per 100 patients years (9 vs. 18, P = 0.02) were lower for S-ICD patients, while inappropriate shock burden was equal (11 vs. 13, P = 0.56). // Conclusion: The earliest experience of the S-ICD demonstrates similar outcomes as contemporary TV-ICD therapy in a matched comparison with predominately single-chamber devices at 3 years follow-up.

Type: Article
Title: Propensity score matched comparison of subcutaneous and transvenous implantable cardioverter-defibrillator therapy in the SIMPLE and EFFORTLESS studies
Location: England
Open access status: An open access version is available from UCL Discovery
DOI: 10.1093/europace/euy083
Publisher version: https://doi.org/10.1093/europace/euy083
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Implantable cardioverter-defibrillator, complications, shocks
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Clinical Science
URI: https://discovery.ucl.ac.uk/id/eprint/10052534
Downloads since deposit
118Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item